Capteur de COT CheckPoint de Sievers
- Manufacturer
- Veolia
- HTTS ref.
- HTS-SA-0293
- HTDS ref.
- CAPTEUR DE COT CHECKPOINT DE SIEVERS
- Division
- Sciences Analytiques

Description
Le capteur de carbone organique total (COT) est une solution rapide pour l’échantillonnage diagnostique mais aussi une solution de dépannage pour les applications microélectroniques, pharmaceutiques et énergétiques.
Une plus grande flexibilité lorsque vous mesurez le carbone organique total à de basses concentrations !
Portable / On-Line
Key features
- Capteur de carbone organique total
- 2 modèles : CheckPoint Pharma et CheckPointe (applications pharmaceutiques et d’eau extra-pure)
- On-line : surveillance continue / Portable : transport aux réseaux d’eau
- Peu d’entretien
- Rapidité remarquable des analyses
- Facile d’utilisation
- Protocoles de validation IQ/OQ/PQ fournis
Technical specifications
- Outputs
- Ethernet (Modbus TCP/IP), USB, plus three 4-20 mA, two alarms,
- months.
- English, Chinese, or Japanese
- Accuracy
- ± 5% at 500 ppb C of sucrose2,3
- Altitude
- 3000 meters
- Humidity
- 90% non-condensing
- download.
- CheckPoint highlights
- Dimensions
- 25,40 cm x 30,48 cm x 15,24 cm
- 24 V supply.
- e
- Dimensions H
- 25.40 cm (10.0 in); W: 30.48 cm (12.0 in); D: 15.24 cm (6.0 in)
- key benefits
- or non-pressurized sources. CheckPointe can also
- Analysis Modes
- On-line (average or timed); grab (with optional Vial Sampling Kit)
- Vial Sampling Kit
- APK 97900
- except as follows
- troubleshooting. The CheckPoint Pharma comes with
- Power Requirements
- 100–240 ±10% VAC, 60 W, 50/60 Hz, or battery
- Low Pressure Samples
- -6.9–55 kPa (-1.0 to 8 psig) (-0.069-0.55 bar) with optional Low Pressure Sampling Kit
- Weight ― CheckPoint
- 2.9 kg (6.4 lb)
- of 0.05 to 1,000 ppb.
- water changes, the CheckPoint sensor makes it
- On-Line Sample Pressure
- 103–690 kPa (15–100 psig) (1.0–6.9 bar)
- Precision (CheckPointe)
- The greater of <1% RSD or 0.05 ppb for on-line measurements;
- options and accessories
- ordering information
- Analysis Time Selectable
- 15 seconds to 8 hours on-line mode; 10 minutes grab mode
- Checkpoint with I/O board
- PRD 97100
- Non-Temperature Corrected
- 0.023 μS/cm (43.5 Mohm-cm) to 150 μS/cm (0.00667 Mohm-cm)
- Linear Range (CheckPointe)
- 0.05–1,000 ppb C
- Pharmacopeia specifications.
- e
- Précision de l’instrument
- ± 5% quand on mesure 500 ppb C de saccharose
- Weight ― Vial Sampling Kit
- 0.6 kg (1.3 lb)
- Gamme linéaire (CheckPointe)
- 0,05 ppb à 1 000 ppb C
- Precision (CheckPoint Pharma)
- The greater of <1% RSD or 0.21 ppb for on-line measurements;
- (UPW) monitoring applications.
- non-ionic organics. High pressures and temperatures
- Required Sample Line Flow Rate
- 60 mL/min (high pressure) or 1 mL/min (low pressure)
- CheckPoint installation diagram
- wall mounting bracket dimensions
- Temperature Corrected to 25 °C
- 0.055 μS/cm (18.24 Mohm-cm) to 113 μS/cm (0.00885 Mohm-cm)
- Linear Range (CheckPoint Pharma)
- 0.21–1,000 ppb C
- Temperature (CheckPointe) Sample
- 10–60 °C (50–140 °F); Ambient: 10–40 °C (50–104 °F)
- Conductivity/Resistivity Accuracy
- ± 2.0%3
- Installation/Overvoltage Category
- II
- Conductivity/Resistivity Precision
- ± 0.5% RSD (20–40 °C or 68–104 °F)
- Dimensions with Vial Sampling Kit H
- 25.40 cm (10.0 in); W: 34.87 cm (13.7 in); D: 15.24 cm (6.0 in)
- Gamme linéaire (CheckPoint Pharma)
- 0,21 ppb à 1 000 ppb C
- One UV lamp (6-month continuous use)
- ARK 35001
- Checkpoint TOC sensor ― basic model
- PRD 97000
- Checkpoint In-line Filter Kit (60 μm)
- APK 97300
- Temperature (CheckPoint Pharma) Sample
- 10–90 °C (50–194 °F); Ambient: 10–55 °C (50–131 °F)3
- Conductivity/Resistivity Range for TOC4
- Max 1.4 μS/cm or Min 0.7 M -cm from CO24
- Ozone Compatibility (CheckPoint Pharma)
- 50 ppb O3; 200 ppb O3 for 2 hours daily
- Industrial Ratings & Safety Certifications
- CE, ETL listed. Conforms to UL Std. 61010-1. Certified to CSA C22.2 No. 61010-1
- anytime with the new CheckPoint Pharma and
- pharmaceutical manufacturers with a cost-effective
- power, and other markets, has a dynamic range
- uous on-line TOC monitoring to detect critical ultrapure
- Conductivity/Resistivity Calibration Stability
- Typically 6 months
- I/O Board ― The optional I/O board has three
- Pharmaceutical Industry Part Number
- and stainless steel sampling tubes and a waste
- 40-mL Certified Sample Vials (Case of 72) HMI 90606
- CheckPoint On-Line/Portable Total Organic Carbon
- and flexible process monitor for process trending,
- Sensor-to-Sensor Matching ― The sensors can be
- • Self-contained pump for easy system suitability and
- pharmaceutical Purified Water (PW) and Water for
- contamination in power feed and loop UPW. Controlling
- Injection (WFI) and semiconductor ultrapure water
- cation conductivity can be difficult if its source is
- programmable analog outputs (select from TOC, raw
- Checkpoint TOC sensor ― basic model PRD 97150
- Low-Pressure Sampling Kit ― This includes Teflon
- Printer HMI 97050
- Versatility and Convenience ― The sensors can be
- e
- Vial Sampling Kit ― The CheckPoint Vial Sampling
- Options
- calibrated to a reference TOC instrument, allowing
- calibration testing
- for rapid diagnostic sampling and troubleshooting.
- throughout the pharmaceutical manufacturing facility
- to 1,000 ppb and can measure TOC in hot or ozonat-
- qualification requirements.
- • The CheckPoint , which is designed to meet the
- tronics manufacturers new problem-solving and
- conductivity, temperature- compensated conductivity
- Checkpoint with I/O board PRD 97160
- ductivity measurement (Direct Conductometric). They
- or Cogeneration Condensate Polish Control) ― The
- provide a reliable low-cost TOC approach useful for
- CheckPoint provides sensitive detection of trace organic
- CheckPoint sensors can be used online for continuous
- problems in real time. The CheckPoint Pharma can be
- maintenance. Calibration is typically stable for six
- • Display screen and documentation available in
- binary input (remote start/stop), and two alarms with
- e
- specifications are met at 40–55 °C (104–131 °F)
- conductivity accuracy is ± 2.4%, and TOC accuracy is ± 6.7% on 500 ppb C of sucrose.
- • The CheckPoint Pharma has a dynamic range of 0.21
- (SOP) to meet installation, operation, and performance
- Measure low-level total organic carbon (TOC) anywhere,
- Pharmaceutical ― The CheckPoint Pharma provides
- The two CheckPoint models have the same specifications
- for multi-point water monitoring, diagnostics, and
- greater sensitivity requirements of the semiconductor,
- diagnostic capabilities. In addition to providing contin-
- (TOC) sensors. Affordable, versatile, and portable, the
- screening potential trouble areas, and diagnosing
- ity, UV oxidation of organics, and a final, post-UV con-
- Power (UPW, Cation Conductivity Control, Makeup,
- monitoring or hand-carried to any point in a water system
- placed at a dedicated point of use or conveniently moved
Applications
- Microélectroniques
- Pharmaceutiques
HTTSystems product sheet
Download the datasheet in our colours and contact details, or read it full screen.
Request a quote
Reference HTS-SA-0293 is carried into your file. You fill in the context, an engineer takes it from there.
Request a quoteTalk to an engineer
A question about installation, consumables or compatibility before pricing? Our engineers answer from Douala.
Ask the questionClose to this one
Same family, or same division when the family is thin.